Instruction Manual
TE
TE-BA-e-1111
1
DIGITAL COATING THICKNESS GAUGE
(F&FN TYPE)
Model: TE 1250-0.1 FN
Table of contents
1. Features
2. Specifications
3. Front panel descriptions
4. Measuring procedure
5. Calibration
6. Battery replacement
7. Correct handling of coating thickness measurement
with external sensors
8. Trouble Shooting
9. Restore factory settings
10. Notes
11. Declaration of conformity
Annotation: It is strongly recommended to calibrate
the new instrument before the first use, as described
in paragraph 5. By doing this, there will be achieved a
much better measurement result right from the start.
1. Features
»This instrument meets the standards of both ISO 2178,
ISO 2360 as well as DIN, ASTM and BS. It is suitable for
the laboratory and for use in “harsh field” conditions.
»The F mode measures the thickness of nonmagnetic
materials, e.g. paint, plastic, porcelain enamel, copper,
zinc, aluminium, chrome, laquer layers etc. These layers
are located on magnetic materials e.g. steel, iron,
nickle etc. It is often used to measure the thickness of
galvanizing layer, laquer layer, porcelain enamel layer,
phosphide layer, copper tile, aluminium tile, some alloy
tile, paper etc.
»The N mode measures the thickness of nonmagnetic
coatings on nonmagnetic materials.
It is used on anodizing, varnish, paint, enamel, plastic
coatings, powder etc. It can be applied on aluminium,
brass, nonmagnetic stainless steel etc.
»Automatically substrate recognition.
»Manual or automatically auto power off to conserve
batteries
»Two measurement modes: - single and continuous
»Data transfer to PC possible
2. Specifications
Display: 4 digits
Range: 0 to 1250 µm/ 0 to 50 mil
Resolution: 0.1µm (0 to 100 µm)
1 µm (over 100 µm)
Accuracy:
- Standard: 3% of the measured value or min. ± 2.5 µm
Is valid within a tolerance range of ± 100 µm around the
individually measured range, if a two-point calibration was
performed within this tolerance range.
- Off-Set Accur Mode: 1% of the measured value
or min. ± 1.0 µm
Is valid within ± 50 µm around the Off-Set Accur point.
PC- interface: with RS-232C interface
Power supply: 4x 1.5V AAA (UM-4) battery
Operating conditions: Temperature: 0 to 50°C
Humidity: <80%
Size: 126 x 65 x 27mm (5.0 x 2.6 x 1.1 inch)
Weight: about 120g (not including batteries)
Accessories: Carrying case
Operation manual
F-sensor
N-sensor
Calibration foils
Base plate (iron)
Base plate (aluminium)
Optional accessories: Cable & software for RS-232C
3. Front panel description
3- 1 Sensors (F and NF type)
3- 2 Display
3- 3 Zero- key
3- 4 Plus- key
3- 5 Minus- key
3- 6 Power key
3- 7 µm/ mil conversion key
3- 8 Battery compartment
3- 9 Jack for RS-232C interface
3-10 S/C key (single/ continuous measuring mode)