Features
Specifications
Front panel description
Measuring procedure
Calibration
Battery replacement
Trouble shooting
Restore factory settings
Notes
Declaration of conformity

Annotation: It is strongly recommended to calibrate
the new instrument before the first use, as
described in paragraph 5. By doing this it
will be achieved a much better measure-
ment result right from the start.

1.

Features

* This instrument meets the standards of both, ISO 2360
and DIN as well as ASTM and BS.
It is suitable for the laboratory and for use in `harsh field`
conditions.
* The N- mode measures the thickness of nonmagnetic
coatings, e.g. anodizing, varnish, paint, enamel, plastic
coatings, powder etc. These layers are located on non-
magnetic metals e.g. aluminium, brass, nonmagnetic
stainless steel etc.
* Automatic substrate recognition
* Manual or automatically “auto power off” to conserve
batteries.
* Two measurement modes: - single and continuous
* Wide measuring range and high resolution
* Date transfer to PC possible

2. Specifications

Display: 4 digits

Range: 0 to 1250 µm/ 0 to 50 mil

(another range may be specified)

Resolution: 0.1 µm (0 to 100 µm)
1 µm (over 100 µm)
Accuracy:
- Standard: 3% of the measured value or min. ± 2.5 µm
Is valid within a tolerance range of ± 100

μm around the

individually measured range, if a two-point calibration was
performed within this tolerance range.

- Off-Set Accur Mode: 1% of the measured value
or min. ± 1.0 µm

Is valid within ± 50

μm around the Off-Set Accur point.


PC- interface: with RS-232C interface

Power supply: 4x 1.5V AAA (UM-4)

Operating conditions:
Temperature: 0 to 50°C
Humidity: < 80%
Size: 126 x 65 x 27 mm (5.0 x 2.6 x 1.1 inch)

Weight: about 81g (not including batteries)

Accessories: carrying case
- Operation manual
According to Models:
1.incorporate F sensor at model TC1250-0.1F
with button F/N
2. incorporate N-sensor at model
TC1250-0.1N with button F/N
3. incorporate FN-sensor at model
TC1250-0.1 FN, Without button F/N and S/C
- Calibration foils (at each model)
- Base plate (aluminium) at model N
- Base plate (iron) at model F
- 2 Base plates (iron and aluminium) for
Model FN, the combination model

Optional accessories: Cable & software for RS-232C

3. Front panel description

Model TC F and TC N Model TC FN

3-1 Sensor: F, N or FN
3-2 Display
3-3 Zero- key
3-4 Plus- key