4
© Intek, Inc. 2013
Revision D
1.3.3 Rheotherm Cooling Water Flow and Fouling (CWFF) Meters
Primary Calibration Accuracy:
Better than ±2% of reading
Flow Range:
2-20 ft/s (extended range optional)
Repeatability:
±0.5% of reading
Operating Temperature:
Electronics: 40 to 120°F (5 to 49°C)
Sensor: 40 to 140°F (5 to 60°C)
Operating Pressure:
60 psi (1000 psi optional)
Storage Temperature:
-20 to 140°F (-29 to 60°C)
Storage Pressure:
60 psig (maximum)
Wetted Surface:
300 Series SS (material options available)
Neoprene, Polyolefin (cable included)
1.3.4 Pressure/Temperature (PT) Probe
Accuracy:
±0.02 inches HgA for pressure
±0.1°F for temperature
Repeatability:
±0.5% of reading
Operating Temperature:
Electronics: 40 to 120°F (5 to 49°C)
Probe: 40 to 300°F (5 to 149°C)
Operating Pressure:
0.5 to 10 inches Hg absolute
15 psia maximum
Storage Temperature:
-20 to 185°F (-29 to 85°C)
Storage Pressure:
15 psig (maximum)
Process Connection:
Ball valve assembly (1½" thread-o-let must be welded to pipe for installation)
Wetted Surface:
300 Series SS
1.3.5 Thermocouple (TC) Temperature Sensors and Arrays
Temperature range:
40-175°F (higher temperature components optional)
Accuracy:
±0.2°F
Repeatability:
±0.1°F
Wetted Surface:
300 Series SS and thermoplastics