MESSAGE CODE
Description/Symptom
PROBABLE CAUSE
ACTION
OUTPUT ERROR
CODES
PT Probe Message Codes
MSG-2
Reserved
MSG-3
Reserved
MSG-4
Reserved
MSG-5 or (CalX)
Calibration Expired
1. Probe Calibration Expired
1. Contact factory for
Recalibration
N/A
Flow Meter Message Codes
MSG-2
Reserved
MSG-3
Reserved
MSG-4
Reserved
MSG-5
Sensor Issue
1. No sensor communication
2. Irregular sensor data
1. Contact factory
-101
Thermocouple Message Codes
Open
TC Circuit Open
1.
Loose connection
2.
Broken wire
1. Check wiring at the electronics
2. Contact factory
-500
Under
TC Value Under Range
1.
Temperature below 32°F
1. Contact factory
-501
Over
TC Value Over Range
1.
Temperature above 212°F
1. Contact factory
-502
Error
General Error
1.
Loose connection
2.
Component failure
1. Check wiring at the electronics
2. Contact factory
-504
MA (4-20mA DAQ) Message Codes
Open
MA Circuit Open
1. Loose connection
2. Broken wire
1. Check wiring at the electronics
2. Contact factory
-500
Under
MA Value Under Range
1. Output below range
1. Contact factory
-501
Over
MA Value Over Range
1. Output above range
2. Contact factory
-502
Error
General Error
1. Loose connection
2. Component failure
1. Check wiring at the electronics
2. Contact factory
-504