beautypg.com

EVCO EK340AJ7 User Manual

Page 3

background image

Every Contr

ol S.r

.l. • EK 340A • Sheet 2/2

The indications showed by the instrument flashes, except the indications “AL1“ and

“AL2“ (they are alternated with the temperature the chamber probe is reading) and

the buzzer utters an intermittent beep.

8

TECHNICAL DATA

8.1

Technical data

Box: self-extinguishing grey.

Size: 72 x 144 x 79 mm (2.83 x 5.66 x 3.11 in).

Installation: panel mounting, panel cut out 67 x 138 mm (2.63 x 5.43 in), with screw

brackets (supplied by the builder).

Frontal protection: IP 54.

Connections: extractable terminal blocks with pitch 7.5 mm (0.29 in) for cables up

to 2.5 mm² (0.38 sq in, power supply and outputs) and with pitch 5 mm (0.19 in) for

cables up to 2.5 mm² (0.38 sq in, input).

Ambient temperature: from 0 to 55 °C (32 to 131 °F, 10 ... 90% of relative humidity

without condensate).

Power supply: 230 Vac, 50/60 Hz, 4 VA (standard) or 115 Vac, 50/60 Hz, 4 VA

(by request).

Alarm buzzer: included.

Measure inputs: 1 (chamber probe) for “J” or “K” thermocouples.

Working range: from 0 to 700 °C (32 to 999 °F) for “J” thermocouple,

from 0 to 999 °C (32 to 999 °F) for “K” thermocouple.

Setpoint range: from 0 to 999 °C (0 to 999 °F).

Resolution: 1 °F with unit of measure in Fahrenheit, 1 °C with unit of measure in

Celsius.

Display: one red LED 3-digit displays 13.2 mm (0.51 in) high, two LED bars (10 red

LED), output status indicators, indicators of the unit of measure of the temperature

showed by the instrument.

Outputs: 3 relays: one 8 A @ 250 Vac relay for top heating group control (NO),

one 8 A @ 250 Vac relay for floor heating group control (NO), one 8 A @ 250 Vac relay

for chamber light control (NO); the maximum current allowed on terminal 26 is 10 A.

cold joint

alarm

first

tempera-

ture

alarm

second

tempera-

ture

alarm

• the chamber probe

plays up

• the connection in-

strument-chamber

probe is wrong

• the temperature the

chamber probe is

reading is outside

the limits allowed by

the working range

of the instrument

there is a defect in the

cold joint of the instru-

ment

the temperature the

chamber probe is

reading is outside the

limit you have set with

parameter AA1

the temperature the

chamber probe is

reading is outside the

limit you have set with

parameter Ab1

• test connection in-

strument-probe

• test the tempera-

ture close to the

probe

turn OFF the power

supply of the instru-

ment: unless the alarm

disappears, you will

have to change the

instrument

test the temperature

close to the probe (look

at parameters AA0,

AA1 and AA4)

test the temperature

close to the probe (look

at parameters Ab0,

Ab1 and Ab4)

• the top output will be

turned OFF

• the floor output will be

turned OFF

no effect

no effect

10

ELECTRICAL CONNECTION

10.1 Electrical connection

(7)

provide the probe with a protection able to protect it against contacts with metal parts or use insulated probes.