ALTANA BYK-mac i COLOR Manual User Manual
Page 55

54
Error
Instrument temperature
Temperature is outside the operating
out of range
temperature range.
Ambient light
Appears when there is too much ambient
light during measurement or when measuring
bright colors (white and light silver) in direct
sunlight. Please refer to chapter 5
“Getting started/Ambient light“
Please clean
Appears when check on cyan tile failed.
Cyan Tile
Please make sure that temperature and
humidity are within specification. Then clean
all standard tiles and perform calibration and
cyan check again.
Please calibrate
Appears when the instrument is not
calibrated and the user wants to take a
reading.
Humidity too high
Humidity is outside the operating
humidity range.
Battery Low
Replace the batteries or recharge the
rechargeable battery pack.
Input required
Data must be input.
Errors and Warning Messages