ALTANA BYK-mac i COLOR Manual User Manual
Page 20

19
Calibrate
Calibration Notes
•
Dirt or dust in the optics area will cause an
inaccurate calibration reading. Therefore, always
use the protective cap, if no readings are taken.
•
The standards and black trap should be cleaned
periodically. For cleaning procedure see section
"Cleaning and Maintenance“. Please make sure
that the standards are not scratched.
•
The cyan tile is to some extend temperature
dependent.Therefore, make sure to perform
calibration and cyan check in laboratory
environment: temperature (20 – 25°C) and
humidity (30 – 60%) controlled.
•
Do not move the instrument while taking a
calibration measurement. If motion is detected,
an error message will be displayed and
calibration is aborted.
•
When moving from cold to warm environment,
there is a danger of condensation. For this
reason, you should wait for an appropriate
amount of time to allow the optical components
to adjust before calibrating and using the unit.