beautypg.com

Holtgreven Ohaus Defender Series Bench Scales User Manual

Page 52

background image

EN-50 7000 Series Indicators

lIb

W001

lIb

PC001

*

-1.000

kg

*

0

0.000

kg

○ ● ○

UNDER ACCEPT OVER

○ ○ ●

UNDER ACCEPT OVER

*

1.000

kg

lIb

P001

lIb

C001

lIb

D001

Zero Check Weighing
Zero check weighing is used when comparing subsequent samples to an initial reference sample.
In this case, the under limit must be a negative value and the over limit must be a positive value.

Place the reference item on the scale and press the

TARE button. Remove the reference sample and

place the item to be compared on the scale to determine if it is within the accept range.

If the item is lighter than the under limit, the yellow UNDER LED will light. If the item is within the
target weight range, the green ACCEPT LED will light. If the item is heavier than the over limit, the

red OVER LED will light.


4.10 Library

When an item is processed on a regular basis, the item’s data may be stored in memory for future use.

The following data is stored for each mode.

Mode

Record ID Name Preset Tare APW Reference Wt. Under Limit Over Limit Level

Weighing

‰

‰

‰

Counting

‰

‰

‰

‰

Percent Weighing

‰

‰

‰

‰

Check Weighing

‰

‰

‰

‰

‰

Dynamic Weighing

‰

‰

‰

‰


NOTES: LIBRARY must be set to ON in the SETUP menu (Section 3.4.18).

Up to 256 library records may be stored.


4.10.1 Storing Library Data
Press the

LIBRARY button to view the next available memory location for the active mode.


The display shows the Record ID with a mode prefix and a unique identification number.

Wxxx = Weighing records
PCxxx = Parts Counting records
Pxxx = Percent Weighing records
Cxxx = Check Weighing records
Dxxx = Dynamic Weighing records







Press the

No button to advance to the next Record ID OR press the Yes button to begin entering the

library data for the displayed Record ID.

The data type is displayed on the first line and the data value is displayed on the second line. Use
the keypad to change the data value. Press the

Yes button to accept the data value and move to the

next data type.