Hardy HI 6600 Series Modular Sensor System User Manual
Page 31

HI 6600 Series Modular Sensor System User Guide
Page | 31
Suggested Minimum Steps When Setting Up the Instrument For the First Time
Choose a specific Weight Processor Module to set up by selecting a channel either from the front panel display
or on the webserver.
Channel
1) Choose channel to set-up
SET UP
(minimum)
1) Select Unit of Measure
2) Input Total Scale Capacity
3) Select Graduation Size
4) Select Decimal Point Position
5) Input Motion Tolerance
Filter
1) Num Averages
2) Waversaver
Calibration
1) Cal Date
2) Cal Method
a. C2 eCal
b. Hard Cal
3) Complete Cal Procedure