10 repairs, 1 checklist for troubleshooting, Repairs 10.1 checklist for troubleshooting – FRITSCH PULVERISETTE 13 premium line User Manual
Page 46: Repairs
10
Repairs
DANGER!
Mains voltage!
– Before beginning with repair work, unplug the mains
plug and protect the device against being unintention-
ally switched back on.
– Indicate repair work with warning signs.
– Repair work may only be performed by specialised per-
sonnel.
– Put safety equipment back into operation after mainte-
nance work.
10.1 Checklist for troubleshooting
Malfunction
Possible cause
Rectify fault
Mill does not start up
No connection to mains
Plug in mains plug
Mains switch off
Switch on mains switch
Protection switch of the grinding
chamber is open
Close the grinding chamber properly
Protection switch of the grinding
stock container is open
Insert the grinding stock container
properly
Thermo switch has triggered
Wait until device has cooled down
Fuse in control circuit
Have fuse replaced by qualified tech-
nician
Disks do not rotate
Gap width at 0
Set the gap width. See Ä Chapter 6
Sample material blocking the disk
Open grinding chamber and remove
sample material
Mill stops running during operation
Overload! Switching off by motor
protection switch (main switch)
Cool off mill, remove grinding stock,
reduce infeed of grinding stock
Grinding chamber overfilled, jam-
ming of grinding stock that is too
large/hard
Open grinding chamber and empty
Mill has poor grinding result
Grinding disk worn down
Reverse direction of rotation or
replace grinding disk. See
on page 23 or Ä Chapter 6 ‘Using the
Repairs
- 46 -