beautypg.com

10 repairs, 1 checklist for troubleshooting, Repairs 10.1 checklist for troubleshooting – FRITSCH PULVERISETTE 13 premium line User Manual

Page 46: Repairs

background image

10

Repairs

DANGER!
Mains voltage!
– Before beginning with repair work, unplug the mains

plug and protect the device against being unintention-

ally switched back on.

– Indicate repair work with warning signs.
– Repair work may only be performed by specialised per-

sonnel.

– Put safety equipment back into operation after mainte-

nance work.

10.1 Checklist for troubleshooting

Malfunction

Possible cause

Rectify fault

Mill does not start up

No connection to mains

Plug in mains plug

Mains switch off

Switch on mains switch

Protection switch of the grinding

chamber is open

Close the grinding chamber properly

Protection switch of the grinding

stock container is open

Insert the grinding stock container

properly

Thermo switch has triggered

Wait until device has cooled down

Fuse in control circuit

Have fuse replaced by qualified tech-

nician

Disks do not rotate

Gap width at 0

Set the gap width. See Ä Chapter 6

‘Using the device’ on page 23

Sample material blocking the disk

Open grinding chamber and remove

sample material

Mill stops running during operation

Overload! Switching off by motor

protection switch (main switch)

Cool off mill, remove grinding stock,

reduce infeed of grinding stock

Grinding chamber overfilled, jam-

ming of grinding stock that is too

large/hard

Open grinding chamber and empty

Mill has poor grinding result

Grinding disk worn down

Reverse direction of rotation or

replace grinding disk. See

Ä

Chapter 6 ‘Using the device’

on page 23 or Ä Chapter 6 ‘Using the

device’ on page 23

Repairs

- 46 -