5.
Lash the harness to a crane or another hoist.
6.
Next lash the harness to the two eyelets provided for this.
7.
Using the crane, bring the disk mill to the desired position. Make
sure the device does not swing while it is being transported.
8.
Remove the harness.
9.
Replace the housing cover (14) and screw it tight.
4.4 Ambient conditions
WARNING!
Mains voltage!
– The device may only be operated indoors.
– The surrounding air may not carry any electrically con-
ductive dust.
– Maximum relative humidity 80% for temperatures up
to 31°C, linearly decreasing down to 50% relative
humidity at 40°C.
n
The room temperature has to stay between 5 - 40°C.
n
Altitudes up to 2000 m
n
Degree of pollution 2 according to IEC 664.
4.5 Electrical connection
Before establishing the connection, compare the voltage and current
values stated on the type plate with the values of the power supply
system to be used.
CAUTION!
Ignoring the values on the type plate may result in damage
to the electrical and mechanical components.
Installation