6.30 On-Demand Data Logging Mode
When configured for On-Demand Data Logging Mode, up to 100 data points
can be stored in memory along with the calculated average and standard
deviation of the stored values. Each data point can be manually entered into
memory.
When the
key is pressed the currently displayed tension reading will
be stored into memory and the memory location (a number from 1 to 100) will
momentarily flash on the display.
Note: The first time the key is pressed the“ M” indicator will flash
on and off on the LCD display. When the memory is full (100 data points in
memory), the “M” indicator will stop flashing and remain on the display.
6.31 Viewing Data Stored In On-Demand Data Logging Mode
To view data stored in Extended Memory press the key.
The following will be displayed when the key is pressed repeatedly.
Each time the key is pressed the displayed value will change.
To Store a Data Point in memory, press the
key.
STORE
Action
Display Will Show
LCD Indicator
Press
R
ECALL
key Maximum
Value MAX
Press
R
ECALL
key Minimum
Value MIN
Press
R
ECALL
key Average
Value
“
A ”
Press
R
ECALL
key
Standard Deviation
“ d ”
Press
R
ECALL
key
Date Point #1
“ 1 ”
Press
R
ECALL
key
Data Point #2
“ 2 ”
● ●
●
● ●
●
● ●
●
Press
R
ECALL
key
Data Point #100
“ 100 ”
16
STORE
STORE
RECALL
RECALL
Press the
key at any time to exit the Recall Mode.
6.32 To Clear All Data Stored In The Extended Memory.
Press the
key and the key simultaneously.
6.40 Continuous Data Logging Memory Mode
When configured for Continuous Data Logging Mode, up to 100 data points can
be stored in memory along with the calculated average, peak and standard
deviation values of the data stored in memory. Each displayed value is auto-
matically entered into memory at a rate equal to the selected Display Update
Rate of 0.5, 1, 2 or 4 seconds.
When the
key is pressed the “M” indicator on the LCD Display will
begin to flash indicating that each displayed value is being automatically
stored in memory. The indicator will continue to flash until the
key is
pressed again to stop the automatic storage or until the memory is full. The
indicator will stop flashing and remain on the display.
17
MAX
MIN
A
100
2
1
Maximum Value
Data Point 100
Data Point #2
Data Point #1
Minimum Value
Average Value
Standard Deviation
POWER
EXIT
STORE
RECALL
To start or stop the automatic storage of each displayed tension value into
memory, press the
key.
STORE
STORE