Service menu options
To access the service options, click Service from the VISCPRO
®
pri-
mary display and select the desired instrument.
Monitor
The Monitor window permits the user to view current operational values
for the CAV instrument. These settings may be helpful to CANNON
®
technical personnel when troubleshooting difficulties with the CAV.
Select Wash viscometer tube to open the Wash viscometer tube
window and select the desired tube.
Click the desired tube tab; then click Begin Wash to initiate the wash
cycle. The Status: indicator will display and update current actions.
Click Done to exit the Wash viscometer tube window.
NOTES
The Wash operation cannot be executed if the tube is currently running a
sample. To terminate the Wash cycle at any time, click Abort Wash.
CAUTION
The Wash viscometer tube function overrides automatic CAV software
operations. Verify that the tubes are clean and dry before initiating
computer-controlled sample testing.