Cannon Instrument CAV-2000 User Manual
Page 18

CANNON
®
Automatic Viscometer Models CAV-2100 and CAV-2200 with VISCPRO
®
Instruction & Operation Manual
Version 2g — May, 2009; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
12
Procedure
1. Make sure that the spin controls
for Tube and Bulb are both set
to “1”:
This corresponds to the left-hand tube in the left-hand bath and the
bottom bulb in the tube.
2. Check the values for the calibration constant(s) as they appear at the
bottom of the window:
The Calibration window
Compare these values to your archive of the calibration constants (if
available) for tube 1 and bulb 1. The values should be identical.
3. If they are not, place your cursor in the appropriate field, delete the
entry, then type the correct values for the constants in the text boxes.
4. Click Update bulb.
5. Use the tube and bulb spin controls
to select the tubes/bulbs for
which you noted calibration constant errors. Input the correct values
for each. Make certain to click Update bulb after you have corrected
C and E calibration values for each bulb BEFORE selecting the next
tube and/or bulb.