beautypg.com

BUCHI Pump Module C-605 User Manual

Page 41

background image

8 Troubleshooting

41

Pump Combinations Operation Manual, Version E

Table 8-1: Error messages of the Pump Controller C-610 (cont.)

Error number

Possible cause

Remedy

08

Invalid data in pump electronics EEPROM.

Contact the BUCHI customer service

09

RS-485 bus voltage too low.

Contact the BUCHI customer service

10

Internal auxiliary voltage not within tolerance
of pump electronics.

Contact the BUCHI customer service

11

Pump electronics fan defective.

Contact the BUCHI customer service

23

EEPROM C-605.

Contact the BUCHI customer service

25

Transfer error to RS485.

Contact the BUCHI customer service

31

The wrong pump is connected (C-605 instead
of C-601).

Replace C-601 with C-605.

Table 8-2: Error messages of the Pump Manager C-615

Error number

Possible cause

Remedy

01

Pump A: Flow rate out of tolerance.

Control backpressure
Contact the BUCHI customer service

02

No answer from Pump A.

Check RJ 485 connections.
Check main power.
Check switch + connections.
Check selector switch at the back side of the
pump module.

03, 14, 25

Communication failure on RS485.

Contact the BUCHI customer service

04

EEPROM pump A.

Contact the BUCHI customer service

05

Pump A: Temperature out of tolerance.

Contact the BUCHI customer service

06

Pump A: Main voltage above tolerance.

Contact the BUCHI customer service

07

Pump A: Main voltage below tolerance.

Contact the BUCHI customer service

08

Pump A: Invalid data in memory.

Contact the BUCHI customer service

09, 10

Failure in pump A.

Contact the BUCHI customer service

11

Cooling fan in Pump A not working.

Contact the BUCHI customer service

12

Pump B: Flow rate out of tolerance.

Contact the BUCHI customer service

13

No answer from Pump B.

Check RJ 485 connections.
Check main power.
Check switch + connections.
Check selector switch at the back side of pump
module.

15

EEPROM pump B.

Contact the BUCHI customer service

16

Pump B: Temperature out of tolerance.

Contact the BUCHI customer service

17

Pump B: Main voltage above tolerance.

Contact the BUCHI customer service

18

Pump B: Main volatage below tolerance.

Contact the BUCHI customer service

19

Pump B: Invalid data in memory.

Contact the BUCHI customer service

20, 21

Failure in pump B.

Contact the BUCHI customer service