BUCHI Pump Module C-605 User Manual
Page 41

8 Troubleshooting
41
Pump Combinations Operation Manual, Version E
Table 8-1: Error messages of the Pump Controller C-610 (cont.)
Error number
Possible cause
Remedy
08
Invalid data in pump electronics EEPROM.
Contact the BUCHI customer service
09
RS-485 bus voltage too low.
Contact the BUCHI customer service
10
Internal auxiliary voltage not within tolerance
of pump electronics.
Contact the BUCHI customer service
11
Pump electronics fan defective.
Contact the BUCHI customer service
23
EEPROM C-605.
Contact the BUCHI customer service
25
Transfer error to RS485.
Contact the BUCHI customer service
31
The wrong pump is connected (C-605 instead
of C-601).
Replace C-601 with C-605.
Table 8-2: Error messages of the Pump Manager C-615
Error number
Possible cause
Remedy
01
Pump A: Flow rate out of tolerance.
Control backpressure
Contact the BUCHI customer service
02
No answer from Pump A.
Check RJ 485 connections.
Check main power.
Check switch + connections.
Check selector switch at the back side of the
pump module.
03, 14, 25
Communication failure on RS485.
Contact the BUCHI customer service
04
EEPROM pump A.
Contact the BUCHI customer service
05
Pump A: Temperature out of tolerance.
Contact the BUCHI customer service
06
Pump A: Main voltage above tolerance.
Contact the BUCHI customer service
07
Pump A: Main voltage below tolerance.
Contact the BUCHI customer service
08
Pump A: Invalid data in memory.
Contact the BUCHI customer service
09, 10
Failure in pump A.
Contact the BUCHI customer service
11
Cooling fan in Pump A not working.
Contact the BUCHI customer service
12
Pump B: Flow rate out of tolerance.
Contact the BUCHI customer service
13
No answer from Pump B.
Check RJ 485 connections.
Check main power.
Check switch + connections.
Check selector switch at the back side of pump
module.
15
EEPROM pump B.
Contact the BUCHI customer service
16
Pump B: Temperature out of tolerance.
Contact the BUCHI customer service
17
Pump B: Main voltage above tolerance.
Contact the BUCHI customer service
18
Pump B: Main volatage below tolerance.
Contact the BUCHI customer service
19
Pump B: Invalid data in memory.
Contact the BUCHI customer service
20, 21
Failure in pump B.
Contact the BUCHI customer service