Fault

Possible cause

Remedy

Optimum distilla­
tion not
achieved

Too much back­evaporation
from the receiving flask (espe­
cially with solvent mixtures)

u

Empty receiving flask and
restart distillation (see Chap­
ter 6.3.2 "Removing the re­
ceiving flask", page 38
,
Chapter 6.1.4 "Fitting the re­
ceiving flask", page 30
 and
Chapter 6.2 "Performing dis­
tillation", page 32).

Other faults in distillation
process (e.g. sudden cooling,
too little heat flow, etc.)

u

Check temperatures of heat­
ing bath and coolant and ad­
just as necessary (see
Chapter 6.1.1 "Preparing the
heating bath", page 27
 and/
or cooler operating instruc­
tions).

u

Lower pressure until distilla­
tion resumes (see vacuum
pump operating instructions).