Help with faults | 8
Operation Manual Rotavapor® R100
43/60
Fault
Possible cause
Remedy
Optimum distilla
tion not
achieved
Too much backevaporation
from the receiving flask (espe
cially with solvent mixtures)
u
Empty receiving flask and
restart distillation (see Chap
ter 6.3.2 "Removing the re
ceiving flask", page 38,
Chapter 6.1.4 "Fitting the re
ceiving flask", page 30 and
Chapter 6.2 "Performing dis
tillation", page 32).
Other faults in distillation
process (e.g. sudden cooling,
too little heat flow, etc.)
u
Check temperatures of heat
ing bath and coolant and ad
just as necessary (see
Chapter 6.1.1 "Preparing the
heating bath", page 27 and/
or cooler operating instruc
tions).
u
Lower pressure until distilla
tion resumes (see vacuum
pump operating instructions).