beautypg.com

6 electrical connections, 7 overview of connections – BUCHI Rotavapo R-100 User Manual

Page 25

background image

BÜCHI Labortechnik AG

Installation | 5

Operation Manual Rotavapor® R­100

25/60

5.6

Electrical connections

u

Make sure that the available power supply matches the rating indicated on the type
plate.

u

Plug the rotary drive unit connecting lead into the socket (2) on the heating bath.

u

Plug the power cord into the socket (1) on the heating bath and then into a mains
power socket.

u

If an extension lead is required, make sure that it is earthed and has a suitable
power rating.

5.7

Overview of connections

The diagram below shows the connection of tubing and electrical leads in a typical ap­
plication of the Rotavapor® R­100 in conjunction with a Recirculating Chiller F­105, a
Vacuum Pump V­100, an Interface I­100 and a Woulff bottle (see also Chapter 3.2.4
"Typical application", page 14).

IN

OUT

The connections to the Recirculating Chiller F­105, the Vacuum Pump V­100 and the
Interface I­100 are described in the particular operating instructions in each case.