6 | Operation

BÜCHI Labortechnik AG

32/60

Operation Manual Rotavapor® R­100

6.2

Performing distillation

 DANGER

Risk of poisoning from inhalation of hazardous vapors

u

Do not inhale any vapors produced during distillation.

u

Ensure that vapors are removed immediately by a suitable fume hood.

u

Only use the device in well ventilated areas.

u

If vapors escape from joints or connections, check the seals concerned and re­
place them if necessary.

u

Do not distill any unknown fluids.

u

Observe the safety data sheets for all fluids used.

 DANGER

Explosion risk from distillation of hazardous materials

u

Do not distill solvents that produce explosive mixtures of gases.

u

Make sure that there is always a sluggishly reactive atmosphere inside the system.

u

Discharge any electrostatic charge by suitable grounding arrangements.

u

Keep sources of ignition away from the apparatus.

 WARNING

Risk of explosion from excessive internal pressure

The evaporating flask and/or the condenser could explode if the internal pressure pro­
duced by evaporation becomes too great.

u

Make sure that the internal pressure inside the glass components is never greater
than atmospheric pressure.

 CAUTION

Risk of skin burns from hot components

u

Do not touch hot components unless wearing suitable protective gloves.

To achieve optimum distillation results, the following procedure is recommended:

u

Adjust the flow rate of the coolant to suit the type of coolant and the cooling capac­
ity. If tap water is used, the flow rate should be at least 40 L/h.

u

Make sure that the coolant temperature is not more than 20 °C.

u

Pour solvent into evaporating flask and fit evaporating flask (see Chapter 6.1.2 "Fit­
ting the evaporating flask", page 28).

u

Make sure that the receiving flask is fitted.

u

Close the glass stopcock.

u

Switch on the rotary drive unit and adjust the rotation speed if necessary.

u

Set the vacuum so that the boiling point of the solvent is 20 °C below the tempera­
ture of the heating bath.