5.4

Fitting the glass stopcock

Materials required:

— Grease Glisseal 40 blue (order number 048197)

u

Connect a supply tubing for solvent (2).

u

Apply a thin coat of Grease Glisseal 40 blue to the glass stopcock (1).

u

Insert glass stopcock into the condenser and turn to the desired position.

The glass stopcock performs various functions depending on its position:

— Glass stopcock handle pointing forwards or backwards: the system is sealed. This

is the standard position for distillation.

— Glass stopcock handle pointing upwards: system is being vented.

— Glass stopcock handle pointing downwards: additional solvent can be fed into the

evaporating flask via a connected feeding tubing.