3 finishing distillation, 1 removing the evaporating flask – BUCHI Rotavapo R-100 User Manual
Page 37
BÜCHI Labortechnik AG
Operation | 6
Operation Manual Rotavapor® R100
37/60
6.3
Finishing distillation
u
u
Vent the Rotavapor.
u
Turn off the rotary drive unit (set the rotation speed to 0 rpm).
u
Switch off the heating bath
u
CAUTION! Risk of skin burns from hot evaporating flask. Check the tempera
ture of the evaporating flask and, if necessary, allow it to cool down or else wear
suitable protective gloves.
u
Remove the evaporating flask (see Chapter 6.3.1 "Removing the evaporating
flask", page 37).
u
Turn off the coolant supply.
u
Remove the receiving flask (see Chapter 6.3.2 "Removing the receiving flask",
page 38).
u
Dry the Rotavapor (see Chapter 7.6 "Removing solvent accumulations", page 41).
u
Clean the Rotavapor and all glass components (see Chapter 7 "Cleaning and ser
vicing", page 39).
6.3.1
Removing the evaporating flask
1
3
4
u
u
Vent the Rotavapor.
u
Turn off the rotary drive unit (set the rotation speed to 0 rpm).
u
CAUTION! Risk of skin burns from hot evaporating flask. Check the tempera
ture of the evaporating flask and, if necessary, allow it to cool down or else wear
suitable protective gloves.
u
Firmly hold the evaporating flask (4) and release the combiclip (3) by screwing it
counterclockwise.
u
Push the clip (2) back from the neck of the evaporating flask.
u
Detach the evaporating flask.
u
If the evaporating flask sticks because of chemical residues, press the lock release
button (1) and twist the flask (4) counterclockwise to remove it.