5 safety features, 1 overheat cut-out, 2 protection against electrical overload – BUCHI Rotavapo R-100 User Manual
Page 19: 3 clips and holders, 4 glassware, 5 optional accessories
BÜCHI Labortechnik AG
Product description | 3
Operation Manual Rotavapor® R100
19/60
3.5
Safety features
3.5.1
Overheat cutout
To protect against overheating, the heating bath has a thermostatic bath temperature
control.
In addition it is fitted with an electronic and a mechanical overheat cutout.
The electronic overheat cutout monitors the temperature limit (the actual temperature
of the bath must not exceed the specified temperature by more than 2 °C for longer
than 2 minutes) the heating rate ( the actual temperature must not increase by more
than 5 °C within 5 seconds) and the function of the temperature sensor. If the elec
tronic overheat cutout trips, the device should be inspected by a BUCHI service tech
nician.
The mechanical overheat cutout consists of a bimetallic thermostat that immediately
cuts off the power supply at high temperatures (above 145 °C). The mechanical over
heat cutout has to be manually reset once the heating bath has cooled down (see
Chapter 8.2.1 "Resetting overheat safety cutout", page 44).
3.5.2
Protection against electrical overload
The heating bath is equipped with fuses to protect against electrical overload.
The rotary drive unit has an electrical overload cutout.
3.5.3
Clips and holders
Combiclips for fixing the evaporating flask and safe release of sticking ground
glass joints
Ball joint clamps for securely fixing the receiving flask
Laboratory stand and holder for fixing glass apparatus
Cap nut for fixing the condenser
3.5.4
Glassware
Highquality borosilicate glass 3.3
GL14 hose barbs for preventing glass breakage
3.5.5
Optional accessories
Safety guard to protect user against glass splinters, solvent splashes or hot heating
fluid in the event of accidents or implosions
Laboratory stand for providing additional fixing for condenser