Ii. getting started, Ii.1 power on, Ii.2 cone spindle selection and setting – Brookfield CAP2000 Viscometer User Manual

Page 7: Getting started, Ii.1, Power on, Ii.2, Cone spindle selection and setting

background image

Brookfield Engineering Labs., Inc.

Page 7

Manual No. M/94-310F

II.

GETTING STARTED

II.1

Power ON

Turn the power ON using the switch located on the rear of the base console.

The display will sequentially show BROOKFIELD, then the MODEL of the viscometer. Af-
ter about ten seconds, the MAIN SCREEN will be displayed, indicating the temperature of the
sample plate (Figure 2).

DEFAULT TEMPERATURES

CAP L Series Viscometer

25.0

°

C

CAP H Series Viscometer

50.0

°

C

MAIN SCREEN

Figure 2

25.0

°

C

SELECTING CGS OR SI UNITS

Pressing the STOP and the RUN keys simultaneously after power ON will enable you to select
either CGS or SI units of measurement for the display and printer. A momentary CGS or SI
will be displayed on the MAIN SCREEN. If a printer is connected, a new header will be
printed showing the change in measurement unit.

II.2

Cone Spindle Selection and Setting

Raise the viscometer handle to its highest position.

The CAP cones have viscosity ranges as shown in Appendix A. After selecting the appropriate
cone for the viscosity range to be utilized, carefully attach the cone to the viscometer as shown:

CONE ATTACHMENT

Figure 3

Solvent Trap

Cone Spindle

Thumb Screw

Viscometer in

"cleaning" position

Viscometer in

measurement position