INTRODUCTION
The CAP Series Viscometers are Cone and Plate geometry high shear rate instruments with
integrated temperature control of the test sample material. The CAP 2000 has variable shear
rate capability over the speed range from 50 - 2000 RPM. The CAP 1000 is a fixed shear rate
viscometer rotating at 750 RPM on 50 Hz and 900 RPM on 60 Hz power supply. The instru-
ments operate by digital setting and display; rotational speed can be automatically timed to
stop. High shear rate viscosity measurements are made over various viscosity ranges depend-
ing upon the model of the viscometer (CAP 1000 or CAP 2000), the cone spindle and the
rotational speed (shear rate). Viscosity is selectively displayed in either poise (P) or Pascal
seconds (Pa•s). Temperature control of sample is possible between either 5
°
C and 75
°
C or
50
°
C and 200
°
C depending on viscometer model.
The CAP Series Viscometers selectably display in either the CGS or SI units (see page 7):
CGS
SI
Comment
Viscosity:
P
Pa•s
0.1 Pa•s = 1 P (= 100 cP)
Shear Rate:
S
EC
-1
S
EC
-1
Speed:
RPM
RPM
Temperature:
°
C
°
C
The CAP Series Viscometers output data to a computer or parallel printer in the CGS and SI
units:
CGS
SI
Comment
Viscosity:
P
Pa•s
0.1 Pa•s = 1 P (= 100 cP)
Full Scale Range (F.S.R.):
%
%
Shear Stress:
Dynes/cm
2
N/m
2
1.0 N•m = 10
7
dyne•cm
Shear Rate:
S
EC
-1
S
EC
-1
Speed:
RPM
RPM
Run Time:
Seconds
Seconds
Temperature:
°
C
°
C
Cone Spindle Number:
No.
No.
Sample Number:
No.
No.
I.1
Components
The following items are included:
Part No.
1. CAP 1000 or CAP 2000 Viscometer
2. Cone Spindle(s) ...................................... CAP-S-0X (X will be shown as a number 1-6)
3. Spindle Case ......................................... CAP-106Y
4. Thumb Screw Wrench ............................ RK-24
5. Solvent Trap ......................................... CAP-118
6. Power Cord: 115V ................................ DVP-65
220V ................................ DVP-66
UK ................................... DE-8
Germany .......................... DE-7
7. Operating Instructions Manual ............... M/94-310