Appendix a, Cone numbers - sample sizes - viscosity ranges – Brookfield CAP2000 Viscometer User Manual
Page 17

Brookfield Engineering Labs., Inc.
Page 17
Manual No. M/94-310F
APPENDIX A - Cone Numbers - Sample Sizes - Viscosity Ranges
CAP 1000 AND 2000 VISCOMETERS
01
13.33N
67
CAP1L
-
100 cP
CAP5L & 1H
-
140 cP
02
13.33N
38
CAP2L
-
210 cP
CAP2H
-
280 cP
03
13.33N
24
CAP3L
-
400 cP
CAP6L & 3H
-
380 cP
04
3.33N
124
CAP4L
-
750 cP
CAP4H
-
730 cP
05
3.33N
67
CAP5L & 1H
-
1400 cP
CAP5H
-
1400 cP
06
3.33N
32
CAP6L & 3H
-
4900 cP
CAP6H
-
4700 cP
N = RPM
Table A1
Cone
Shear Rate
Sample Size
Calibrating Fluid
Calibrating Fluid
Number
(sec-1)
(micro liters)
Low Temp Cap at 25
°
C
High Temp Cap at 60
°
C
Part No.
*Viscosity
Part No.
*Viscosity
Version 1.0
*Approximate Value
CAP 2000 Viscosity Ranges
Cone
Range
(Poise)
Coefficient
750 RPM
900 RPM
Cone
Number
1
Table A2
Cone
Number
1
Table A3
Cone
Number
1
Cone Angle
Cone Radius
(degrees)
(cm)
Table A4
1
Number is marked on the cone shaft
CAP 1000 VISCOSITY RANGES (POISE)
01
0.25
-
2.5
0.21
-
2.08
02
0.5
-
5.0
0.42
-
4.17
03
1.0
-
10.0
0.83
-
8.33
04
2.0
-
20.0
1.67
-
16.67
05
4.0
-
40.0
3.33
-
33.33
06
10.0
-
100.0
8.33
-
83.33
01
0.09
-
37.5
1,875
02
0.19
-
75.0
3,750
03
0.38
-
150.0
7,500
04
0.75
-
300.0
15,000
05
1.50
-
600.0
30,000
06
3.75
-
1,500.0
75,000
01
0.45
1.511
02
0.45
1.200
03
0.45
0.953
04
1.8
1.200
05
1.8
0.953
06
1.8
0.702