beautypg.com

Iii.1 accuracy of measurement, Iii.1, Accuracy of measurement – Brookfield CAP2000 Viscometer User Manual

Page 13

background image

Brookfield Engineering Labs., Inc.

Page 13

Manual No. M/94-310F

III.1

Accuracy of Measurement

Table 3.1 indicates the accuracy of the viscosity measurement which depends on both the rota-
tional speed of the cone and the percent of full scale range (in Poise) at which viscosity is
measured. The accuracy for viscosity data provided by CAP Viscometers is the indicated per-
centage of the full scale range. See Appendix A for information on how to determine Full Scale
Viscosity Range (FSR).

ACCURACY OF VISCOSITY MEASUREMENT

Table 3.1

CAP 1000L AND CAP 2000L VISCOMETERS

FULLSCALE VISCOSITY RANGE ACCURACY

Cone

750 RPM

900RPM

Over 900 RPM to 2000 RPM

01

±

2.0%

±

2.0%

02

±

2.0%

±

2.0%

03

±

2.0%

±

2.0%

04

±

2.0%

±

4.0%

05

±

2.0%

±

4.0%

06

±

4.0%

±

10.0%

Varies depending on the

thermal conductivity of the

sample being measured

FULLSCALE VISCOSITY RANGE ACCURACY

Cone

750 RPM

900RPM

Over 900 RPM to 2000 RPM

10 to 100% FSR

50% FSR >50% FSR

FSR = Full Scale Range

01

±

2.0%

±

2.0%

±

4.0%

02

±

2.0%

±

2.0%

±

4.0%

03

±

2.0%

±

2.0%

±

4.0%

04

±

3.0%

±

3.0%

±

6.0%

05

±

4.0%

±

4.0%

±

8.0%

06

±

5.0%

±

5.0%

±

10.0%

Varies depending on the

thermal conductivity of the

sample being measured

Viscometer

Temperature Control Accuracy

CAP1000L & H

±

0.2

°

C

CAP2000L & H

±

0.2

°

C

Note: For CAP2000L Viscometers, the accuracy stated at 750 RPM and 900 RPM applies

over the speed range 50 RPM to 750 RPM and 50 RPM to 900 RPM.

This manual is related to the following products: