Upon pressing the CONE key to accept the last digit of the SMC value, the viscometer displays
the message:

Figure B8

SRC (00) = 00.0000

Using the CONE and INDEX keys, enter the Shear Rate Constant (SRC). Refer to
Appendix A.

Note: All Zeros are not allowed.

Upon pressing the CONE key to accept the last digit of the SRC value, the viscometer displays
the message:

Figure B9

SET TEMP 25.0

°

C

Using the TEMP and INDEX keys, enter the calibration temperature.

Upon pressing the TEMP key to accept the last digit of the calibration temperature, the viscom-
eter displays the message:

Figure B10

FLUID = 00000 cP

Using the CONE and INDEX keys, enter the viscosity of the standard fluid in centipoise
(mPa

.

s).

NOTE: Calibration is performed using CENTIPOISE (mPa

.

s) values, NOT POISE values,

for the Viscosity Standard Fluid.

Viscosity is entered as a five (5) digit number referencing from RIGHT to LEFT.

Example:

Viscosity of 433 centipoise (mPa

.

s). Enter 00433.

Viscosity of 12,500 centipoise (mPa

.

s). Enter 12500.

Note: If an incorrect value is entered for the Viscosity Standard Fluid, press
the STOP key and begin over at Step 1.

Zero viscosity is not allowed.