Figure B4

TEMP. = 25.0

°

C

Verify that the calibrating temperature, as shown in the display, is correct and stable. If the
value is incorrect, press STOP, return to Step 1 and begin again.

Press the RUN key to calibrate the cone. The screen will display:

Figure B5

CALIBRATING ( _ _ )

NOTE:

The 2 digits in parentheses identify the cone that is being calibrated.

When calibration is completed, rotation will stop and the screen will display:

Figure B6

CAL COMPLETE

Pressing the PRINT key sends the calibration data to the printer port (Figure B14) and
returns the CAP to the main screen.

6.

Custom Cones

Custom cone spindles are available from Brookfield Engineering. For further information,
contact Technical Sales, Brookfield USA. The following information relates to calibrating
custom cone spindles.

Pressing the CONE key after completing Step 4, the viscometer displays the message:

Figure B7

SMC (00) = 000000

Using the CONE and INDEX keys, enter the Spindle Multiplier Constant (SMC). Refer to
Appendix A.

Note: All Zeros are not accepted.