Brookfield CAP2000 Viscometer User Manual
Page 22
Brookfield Engineering Labs., Inc.
Page 22
Manual No. M/94-310F
Pressing the CONE key after entering the fifth digit displays the message:
Figure B4
TEMP. = 25.0
°
C
Verify that the calibrating temperature, as shown in the display, is correct and stable. If the
value is incorrect, press STOP, return to Step 1 and begin again.
Press the RUN key to calibrate the cone. The screen will display:
Figure B5
CALIBRATING ( _ _ )
NOTE:
The 2 digits in parentheses identify the cone that is being calibrated.
When calibration is completed, rotation will stop and the screen will display:
Figure B6
CAL COMPLETE
Pressing the PRINT key sends the calibration data to the printer port (Figure B14) and
returns the CAP to the main screen.
6.
Custom Cones
Custom cone spindles are available from Brookfield Engineering. For further information,
contact Technical Sales, Brookfield USA. The following information relates to calibrating
custom cone spindles.
Pressing the CONE key after completing Step 4, the viscometer displays the message:
Figure B7
SMC (00) = 000000
Using the CONE and INDEX keys, enter the Spindle Multiplier Constant (SMC). Refer to
Appendix A.
Note: All Zeros are not accepted.