Brookfield CAP2000 Viscometer User Manual
Page 24
Brookfield Engineering Labs., Inc.
Page 24
Manual No. M/94-310F
Figure B14
CONE (00) CALIBRATION
VISCOSITY
FSR
TEMP
S.STRESS
S.RATE
SPEED
TIMER
CONE
SAMPLE
(POISE)
(%)
(Deg C)
(D/CM
2
)
(1/sec)
(RPM)
(SEC)
No.
No.
Operator:
Date:
Model/Serial#:
Fluid:
Figure B11
TEMP. = 25.0
°
C
Verify the calibrating temperature is correct and stable. If the value is incorrect, press STOP,
return to Step 1 and begin again.
Press the RUN key to calibrate the cone. The screen will display:
Figure B12
CALIBRATING (00)
When calibration is completed, rotation will stop and the screen will display:
Figure B13
CAL COMPLETE
Pressing the PRINT key sends the calibrating parameters to the printer port (Figure B14) and
returns the CAP to the main screen.