Brookfield Engineering Labs., Inc.

Page 10

Manual No. M/94-310F

again and the full scale viscosity is shown (Figure 9). Press RPM once more and the display
will revert to the MAIN SCREEN (Figure 2).

SR: 10000 /sec

Figure 8

100% = 2.500 P

Figure 9

II.5

Temperature Control Setting

Press the TEMP key and the screen will display the current temperature setting. The default
temperature on start-up is 25.0

°

C on low temperature models (Figure 10) and 50

°

C on high

temperature models.

SET TEMP 25.0

°

C

Figure 10

The temperature ranges are:

Low temperature: 5.0

°

C to 75.0

°

C

High temperature: 50.0

°

C to 200.0

°

C

Note: The temperature can be set in increments of 0.1

°

C.

Using the TEMP and INDEX keys, enter the temperature for controlling the sample during
measurement. After the digit in the tenths column is selected and the TEMP key is pressed, the
sequence is completed and the control temperature is set. The display will revert to the main
display screen and show the viscometer going to the control temperature setting.

II.6

Run Time Setting

Press the TIMER key and the screen will be displayed as shown in Figure 11.

SET RUN TIME 020

Figure 11

The default time for the run time setting is 15 seconds.

The timer range is 10 to 999 seconds.