Before you install the Welch Allyn Connex CSK software do the following:
•
If connecting to a CardioPerfect Workstation, do the following:
•
Verify that the CardioPerfect Workstation software and database have both been
upgraded to version 1.6.0 or higher. If necessary, upgrade the software and
database now.
•
When using a stand-alone CardioPerfect Workstation system, please install the
server software on the stand-alone CardioPerfect Workstation system
•
When using a client/server configuration, please install the server software on the
CardioPerfect Workstation database server before installing the clients.
•
Verify that the connectivity settings in the software are properly configured. For
details, see the electrocardiograph directions for use.
•
If using a CP 200 electrocardiograph in a wireless configuration, do the following.
•
Verify that the electrocardiograph software product version is 2.4.0 or later.
•
Verify that the Wireless Settings item is available on the Connectivity Settings
menu. If not, a hardware upgrade is needed. Contact your Welch Allyn service
representative.
•
Check the electrocardiograph’s network and wireless settings. Verify that these
settings match the access point settings.
For details, see the electrocardiograph directions for use.
•
Verify that you have administrator rights for the CardioPerfect Workstation software
and for the PC.
•
If your operating system offers User Access Control (UAC), disable the feature before
installation.
•
Close all programs before starting the installation.
•
If “Device driver signing” appears, select either “Yes” or “Continue.”
To install the Welch Allyn Connex CSK software
1.
Insert the Connex CSK CD. This screen appears: