About quality feedback

The spirometer provides two kinds of quality feedback: effort-quality messages and test

quality grades, as described in the following sections.

About effort-quality messages

One of the following effort-quality messages appears on the screen after each effort is

completed. These messages indicate whether an effort was acceptable and

reproducible, and if not, what the patient needs to do differently.

The term “match” here means “variation” or “difference with respect to best test.”

Effort-quality message

Criteria

Don’t hesitate

Back-extrapolated volume > 150 ml or 5%, whichever is greater.

Blast out faster

PEF time > 120 ms.

Blow out longer

FET < 6.0 seconds, and end-of-test volume > 100 ml (invalid FEV6).

Blast out harder

PEF is not reproducible (match > 1.0 L/s).

Deeper breath

FEV6 match > 150 ml FVC may be substituted for FEV6.

Good effort

Effort meets above criteria.

Good test session

Two acceptable efforts match.

About test-quality grades

Another type of feedback is the test-quality grade, as described in the following table. If

Quality grades is enabled in the FVC report settings, a grade appears on printed reports

and also displays on screen when you view the values or interpretation of a completed

test.

To enable or disable this setting, see Viewing or changing the spirometry settings.

Test-

quality

grade

Number of acceptable efforts

Reproducibility

A

2 or more

Largest two FEV1 values match ≤ 100 mL.
Largest two FVC values match ≤ 100 mL.

B

2 or more

Largest two FEV1 values match > 100 and ≤ 150 mL.

C

2 or more

Largest two FEV1 values match > 150 and ≤ 200 mL.

D

1 or more

Largest two FEV1 values match > 200 mL.

F

None

56 Predictive Norms, etc.