About quality feedback, About effort-quality messages, About test-quality grades – Welch Allyn CP 150 spirometry option - User Manual User Manual
Page 60

About quality feedback
The spirometer provides two kinds of quality feedback: effort-quality messages and test
quality grades, as described in the following sections.
About effort-quality messages
One of the following effort-quality messages appears on the screen after each effort is
completed. These messages indicate whether an effort was acceptable and
reproducible, and if not, what the patient needs to do differently.
The term “match” here means “variation” or “difference with respect to best test.”
Effort-quality message
Criteria
Don’t hesitate
Back-extrapolated volume > 150 ml or 5%, whichever is greater.
Blast out faster
PEF time > 120 ms.
Blow out longer
FET < 6.0 seconds, and end-of-test volume > 100 ml (invalid FEV6).
Blast out harder
PEF is not reproducible (match > 1.0 L/s).
Deeper breath
FEV6 match > 150 ml FVC may be substituted for FEV6.
Good effort
Effort meets above criteria.
Good test session
Two acceptable efforts match.
About test-quality grades
Another type of feedback is the test-quality grade, as described in the following table. If
Quality grades is enabled in the FVC report settings, a grade appears on printed reports
and also displays on screen when you view the values or interpretation of a completed
test.
To enable or disable this setting, see Viewing or changing the spirometry settings.
Test-
quality
grade
Number of acceptable efforts
Reproducibility
A
2 or more
Largest two FEV1 values match ≤ 100 mL.
Largest two FVC values match ≤ 100 mL.
B
2 or more
Largest two FEV1 values match > 100 and ≤ 150 mL.
C
2 or more
Largest two FEV1 values match > 150 and ≤ 200 mL.
D
1 or more
Largest two FEV1 values match > 200 mL.
F
None
56 Predictive Norms, etc.