Performing a spirometry test using the search tab – Welch Allyn CP 150 spirometry option - User Manual User Manual
Page 37

14. Touch Done when you have completed the pre-tests.
If the Auto Save setting is turned off, touch the Yes button and touch Save to save
the test. Select one of the following locations:
•
Local (internal memory)
•
USB mass storage device (Any tests that you save to a USB mass storage
device can be retrieved only from a CardioPerfect workstation.)
•
Workstation
•
Remote file location
Performing a spirometry test using the Search tab
CAUTION Patient data is not saved until the spirometry test is completed.
Note
The spirometry configuration settings can be changed in the Settings tab.
The following settings may appear differently if the default settings have
been modified.
1. Touch the
(Perform new Forced Vital Capacity) button. The Summary tab
appears.
2. Search for patient.
The Search tab gives you access to patient data in the Saved tests directory or in
a connected database (CardioPerfect workstation or EMR).
•
Touch the Search tab.
•
Enter the Patient ID or Last name.
•
Touch the OK button.
•
Touch the Search button.
•
Touch the Patient name , Patient ID, or Birth date in the Patient name/Patient ID
field.
•
Touch the Select button.
•
Touch the (Next) button to review or edit patient information.
•
Touch the (Next) button again.
3. Touch the View or Incentive button to select the display information that you want
to view during the test.
Directions for use
Spirometry tests 33