Inspection and checklist report, Every six months – Welch Allyn 1500 Patient Monitor v.1.3.X - User Manual User Manual
Page 98

94
Maintenance
Welch Allyn 1500 Patient Monitor
Inspection and checklist report
In accordance with the maintenance interval detailed previously, the following check list
should be copied and followed.
Monitor Serial Number: _________________________________
Every six months
Inspection
Result
Checked
General examination
Visual inspection of the
monitor.
Monitor casing not broken or
cracked.
Visual inspection of the
LCD.
LCD screen undamaged.
Visual inspection of all
cable assemblies and
sensors and accessories.
Electrode cable sheathing and
connectors undamaged.
No kinks, abrasion or wear in any
cable assembly.
All transducers and accessories
are within their expiratory date.
Plug and socket
connectors.
Input/output connectors
undamaged.
Button check
Buttons work.
Speaker check.
Switch the monitor on by
pressing the On button.
Switch-on beeps sounded. The
standard screen is displayed.
LED check
Mains LED on when mains
connected. Battery LED on when
battery charging.
Alarm Check
Connect SpO
2
sensor to
volunteer.
Measurement within range.
Set SpO
2
limit to lowest
setting.
Measurement out of range and
visual and audible alarm
activated.
Recurrent test
Confirm the date of last
factory inspections and
test.
If the monitor is due for a yearly
test, have a qualified technician
perform the test.
Date of Inspection:
Inspector: