Messages – Welch Allyn 1500 Patient Monitor v.1.3.X - User Manual User Manual
Page 75

Directions for use
Monitoring and Measurements
71
SpO
2
messages
Message
Possible cause
Suggested action
SpO
2
– check sensor Defective SpO
2
sensor
Replace the sensor.
Incorrect settings in the monitor Check the monitor settings.
SpO
2
– check sensor
placement
Poor sensor contact or the
sensor has fallen off
Check the contact between the sensor and
the patient.
Sensor is disturbed by ambient
light
Cover the sensor.
Sensor defective (red light on
the sensor is not lit)
Replace the sensor.
SpO
2
Low Perfusion
1
1.
When the monitor has a Masimo module, the message SpO2 low perfusion only occurs when the perfusion index
feature is enabled.
Sensor not properly applied
Check the sensor and reapply.
Fingernail varnish on the finger
Remove fingernail varnish.
Thick skin
Change finger.
Sensor failed
Change sensor.
Weak Signal
see SpO
2
Low Perfusion above
see SpO
2
Low Perfusion above.
SpO
2
unplugged
SpO
2
not connected to the
monitor
Connect sensor.
SpO
2
artifact
Patient has moved
Calm the patient.
Hemodynamic interference
Apply sensor to another extremity.
Too thin skin
Apply sensor to a larger finger.
SpO
2
< [lower limit]
SpO
2
too low
Check the patient and alarm limits.
SpO
2
> [upper limit]
SpO
2
too high
Check the patient and alarm limits.
SpO2 low signal IQ
2
2.
The SpO2 low signal IQ message is only given when the monitor has a Masimo module and only occurs if Signal IQ
waveform feature is enabled.
Poor quality signal
Check the sensor and reapply. Move sensor
to obtain a better quality signal.
PR < [lower limit]
Pulse rate too low
Check the patient and alarm limits.
PR > [upper limit]
Pulse rate too high
Check the patient and alarm limits.