Common Analysis problems:

Problem Solution

Sample Burns

Reduce temperature
Use glass fiberfilter on the top of the sample
Reduce sample quantity and distribute it uniformly

Drying takes too long

Increase temperature
Reduce sample mass

A sample is liquid or paste

Use glass fiberfilter

A sample does not consist of enough
volatile matters

Enlarge a sample


Troubleshooting:

Problem Possible

cause Solution

Initial heating “Td” temperature
exceeds 105°C,

The temperature sensor is damaged.

Contact the manufacture.

Initial heating “Td” temperature does
not reach 105°C within 3 minutes,
the halogens do not light up.

The heater is damaged.

Contact the manufacture.

„Tare range exceeded”

Tare key pressed during zero
indication

Balance indications must be different
than zero

„Zeroing range exceeded”

Allowed zeroing range was exceeded

Remove the load from the pan

„Weighing range exceeded”

Maximum weighing capacity
(Max +9e) was exceeded

Reduce the load

„Measuring range exceeded (+)”

Upper limit of analog-digital
transducer measuring range was
exceeded

Remove the load from the pan

„Measuring range exceeded (-)”

Lower limit of analog-digital
transducer measuring range was
exceeded

Check pan for proper installation.
Make sure all parts of the pan are
installed/