Chapter 14: drying options – Torbal ATS120 User Manual

Page 24

background image

- 23 -

Instruction Manual – ATS / BTS Moisture Analyzers


In addition to the” Drying Chart” and “Temperature Correction” the analyzer is equipped with “Averaging” and
“Transmission” options.

Averaging – when enabled the analyzer will collect results of performed analysis and calculate an average. The average
will be rolling until the user erases the collected data by selecting “Erasing” under [Drying Options / Averaging /
Erasing].

Transmission – when enabled the analyzer transmits the result of each sample taken to a PC or a Printer (Via USB or
RS232) during an analysis. The result can be sent in “g” (mass only), % (percent only) or both “g & %”.Transmission
should be enabled in “g” (mass only) when collecting data for detailed statistical analysis, or when using the analyzer with
the TORBAL ProMass software application.

OFF

Transmission is OFF sampling results are not be transmitted to a printer or a PC.

g

The analyzer sends the mass result of the sample taken.

%

The analyzer sends the percent (moisture result) of the sample taken.

% + g

The analyzer sends both the percent (moistures result) as well as mass result of the sample taken.

1.

Press the Menu key and use the navigation keys to select option 3 (Drying Options).

2.

Use the navigation keys to select option 1 (Average) or 2(Transmission)

3.

Enable the options by switching Status form OFF to ON.

Chapter 14: Drying Options