6
Auto-centering pan (AX-MC1000PAN) installation procedure
Caution: During installation, do not apply excessive force to the balance.
z
When used as a mass comparator, the balance can achieve even more precise weighing,
by using the auto-centering pan (sold separately), which reduces eccentric loading errors.
1. At step 2 of the glass breeze break installation procedure, secure the breeze break provided with
the AX-MC1000PAN option using the breeze break securing screws provided with the option. At
this time make sure that the side of the breeze break with the screws is placed at the back.
Attach the pan support receptors to the shock absorbers.
2. Place the auto-centering pan guide on the pan support receptors, and then place the
auto-centering pan on the guide.
Note
When replacing the weighing pan
with the auto-centering pan, be sure
to calibrate the balance before
weighing.
Refer to “7. CALIBRATION”.
Breeze break
securing screw
(2 pieces)
Option
AX-MC1000PAN
breeze break
Auto-centering pan guide
Auto-centering pan
Pan support
receptors
(4 pieces)
Side with screws