Extended function – Rice Lake MC Series Mass Comparator, A&D Weighing - MC-6100 User Manual
Page 53

51
14. EXTENDED FUNCTION
The MC series balance has several extended functions equipped for special applications or to
troubleshoot when using the standard functions.
When the "Filter ( f1l)” is set to “ 0 “, extra items are available as shown below.
Class
Item and Parameter
Description
0
1
2
3
4
5
6
f1-b
Averaging range for
the first moving
average
7
Small
Large
When the fluctuation of a weight value is
within the averaging range, the averaging
operation starts to stabilize the value.
When the fluctuation is small, for example,
when weighing or filling a small amount of
sample, the averaging operation is always
performed and the response rate may be
slow. Under such a condition, change the
parameter.
Refer to “Averaging range ( f1-b ) and
averaging time ( f1-t )”
0 No averaging
1 0.5 second
2
1 second
3 1.5 seconds
4 2.0 seconds
5 2.5 seconds
6 3.2 seconds
7 4.8 seconds
f1-t
Averaging time for the
first moving average
8 6.6 seconds
When the fluctuation of a weight value is
within the averaging range, the averaging
operation starts. Once the duration of
averaging operations reaches the averaging
time, moving averaging starts.
This parameter sets the time to start moving
averaging.
Refer to “Averaging range ( f1-b ) and
averaging time ( f1-t )”
0
1
2
3
4
5
f2-b
Averaging range for
the second moving
average
6
Small
Large
Refer to “Filter depending on differences in
the amount to deliver powdery and liquid
material”.
0 No averaging
1 0.5 second
2 1 second
3 1.5 seconds
4 2.0 seconds
5
2.5 seconds
erfnc
Extended
function
f2-t
Averaging time for the
second moving
average
6 3.2 seconds