2 glp report – Rice Lake MC Series Mass Comparator, A&D Weighing - MC-6100 User Manual
Page 42

40
10-2 GLP Report
Set the following parameters to output the report.
z
To print the report, set the “GLP output ( info )” parameter to “1 ”, the “Data output pause
( pU5e )” parameter to “1 ”, and use MODE 3 of the AD-8121B. For details on using the printer,
refer to “12-1 Connection to the AD-8121B Printer”.
z
To output the report to a personal computer using the RS-232C interface, set the “GLP output
( info )” parameter to “ 2 ”.
z
If the time and date are not correct, set the correct time and date in “Clock ( Cl adj )” of the
function table.
Calibration report using the internal mass
When the setting is “info 1 ”:
When the setting is “info 2 ”:
~~~~~~~~~~~
A
~
&
~
D < T E R M >
MO DE L
~~~~
MC-1 00 0< T E RM >
S/ N
~~~~~
012 34 567 < TE R M >
ID
~~~~~~~
AB CDE FG < T E R M >
DA TE < T E R M >
~~~~~~
2 00 9/12 /31< T E RM >
TI ME < T E R M >
~~~~~~~~
1 2:3 4: 56< T E RM >
CA LI BRA TE D(IN T. )< T E R M>
SI GN ATU RE < T E R M >
-- - - - - - - - -- - - - - - < T ER M >
A & D
MODEL MC-1000
S/N 01234567
ID ABCDEFG
DATE 2009/12/31
TIME 12:34:56
CALIBRATED(INT.)
SIGNATURE
- - - - - - - -
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Signature
AD-8121 format
General data format
CR LF
CR
Space, ASCII 20h
Terminator, , or
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
CR
LF