57
Display Error code
Description
(Blinking)
Memory full
The amount of weighing data in memory has reached
the maximum capacity.
Delete the data in memory to store new data. For
details, refer to the GX series instruction manual, “12.
DATA MEMORY”.
(Illuminated)
Memory full
The amount of calibration or calibration test data in
memory has reached the maximum capacity (50 sets).
The data in memory will be deleted automatically to
store new data. For details, refer to the GX series
instruction manual, “12. DATA MEMORY”.
Memory type error
The type of memory set in the function table and the
type of data stored are different.
For details, refer to the GX series instruction manual,
“12. DATA MEMORY”.
EC, E00
Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
EC, E01
Undefined command error
An undefined command was received.
Confirm the command.
EC, E02
Not ready
A received command can not be processed.
e.g. The balance received a Q command, but not in the
weighing mode.
e.g. The balance received a Q command while
processing a RE-ZERO command.
Adjust the delay time to transmit a command.
EC, E03
Timeout error
If the timeout parameter is set to “ t-Up1 ”, the balance
did not receive the next character of a command within
the time limit of one second.
Confirm the communication.
EC, E04
Excess characters error
The balance received excessive characters in a
command.
Confirm the command.