12. CONNECTION TO PERIPHERAL EQUIPMENT
12-1 Connection to the AD-8121B Printer
Set the following parameters to use the AD-8121B printer.
AD-8121B
Class Item
Factory
setting
Mode 1
Mode 2
Mode 3
prt
Data output mode
0
0, 1, 2, 4, 5
3
0, 1, 2, 4, 5
ap-p
Auto print polarity
0
ap-b
Auto print difference
1
**1
Not
applicable
*1
d-no
Data number output
0 0
0 0, 1
5-td
Time/date output
0 0
0 0,1, 2, 3
5-id
ID number output
0 0
0 0,1
pU5e
Data output pause
0 0
0 0,1 *2
dout
Data output
at-f
Auto feed
0 0
0 0,1
bp5
Baud rate
2
2 2 2
ptpr
Data bit, parity bit
0
0 0 0
Crlf
Terminator
0
0 0 0
type
Data format
0
0 0 1
5if
Serial
interface
Ct5
CTS, RTS control
0
0 0 0
*1 Set the parameter when the data output mode is set to the auto print mode ( prt 1 or prt 2 ).
*2 When multiple lines are to be printed, set the parameter to 1.
AD-8121B settings
Mode
AD-8121B DIP switches
Description
Mode 1
Prints upon data receipt
Standard mode, statistical calculation mode
Mode 2
Prints using the AD-8121B DATA key or the AD-8121B
built-in timer
Standard mode, interval mode, chart mode
Mode 3
Prints upon data receipt
Dump print mode
DIP switch 3: Handling unstable data
DIP switch 4: Data input specification
ON = To print unstable data
ON = Use the current loop
OFF = Not to print unstable data
OFF = Use the RS-232C
With data 0, the weight data can be output to the printer.
With data 2, the weight data stored in memory can be output to the printer.
With data 3, the calibration report stored in memory can be output to the printer.
Refer to “10. ID NUMBER AND GLP REPORT” for print samples.