23

14. Remove the bracket mounted on the drive unit that is attached to cable carrier.

Two #6 screws hold the bracket to the unit.

15. Remove the cable clip holding the power cable.

16. Remove the bracket on the opposite side of the unit that ties the drive unit to the

turntable top. This temporary bracket holds the unit in place for shipping.

17. Turn the brake knob to release the drive unit and allow it to move toward center of

table.

When the turntable top is in place, use the brake knob to adjust the EUT to the

center of rotation (middle of the quiet zone) by sliding the mast assembly back

and forth.

18. Remove the two 1/4–20 hex head screws that hold the wood top support bracket,

and then remove the bracket.

19. Slide the drive unit carrier out.

20. Reinstall in reverse order.

The brake knob must be in the upper position to allow the drive unit to slide

onto the rail system. Verify all hardware is secured.

Bore Sight and Leveling

If the MAPS unit was ordered with multiple masts, you must bore sight

each mast.

B

ORE

S

IGHT

LASER WARNING. Denotes a laser is part of the operating system of the

device.

Bore sight of the MAPS is critical to the accuracy of measurements, and is the

most important step of the installation process. Take the time to verify all