beautypg.com

ETS-Lindgren 2115 Multi-Axis Positioning Systems (MAPS) User Manual

Page 20

background image

20

|

Installation

Reference Point

If installing the MAPS in an existing chamber: Remove the absorber from

the floor and lower wall areas prior to installation to avoid damage to the

absorber.

1. Locate the reference point. It

is generally located along the

bore sight axis of the range

antenna. See Bore Sight and

Leveling on page 23 for

additional information

regarding bore sight.

2. With permanent marker, place

an X on the floor of the

chamber at the reference

point.

3. Draw a 47-in (1.19-m) diameter circle to represent the turntable perimeter.

The diameter is larger than the actual perimeter of the circular anchor plates

for the turntable, and should only be used as a guide in centering the turntable

portion of the MAPS.

System Installation

Fiber optic cables must be connected correctly for motor base

function. Before removing fiber optic cables from the

motor base, label the replacement locations for accurate

reconnection.

4. Remove the wood deck. See assembly drawing 110073 located in the back pocket of

the manual for details.