Good Shielding Practices
| 7
1.0 Good Shielding Practices
To retain the RF shielding integrity of the chamber, follow these good shielding
practices:
•
Periodically clean and lubricate doors.
•
When mounting screw-retaining items within the enclosure, do not
penetrate through both shields of a modular panel system. Screws
must not penetrate any portion of the screen system.
•
Do not use framing screws to fasten items within the enclosure.
•
When a hole is bored for a new penetration, remove all burrs. Clean
the perimeter of the hole and the penetration with a cleaning agent
before installing the new penetration.
•
Nothing must penetrate the enclosure wall, floor, or ceiling without use
of a proper ETS-Lindgren designated penetration.
•
Bolts fitted to the modular and pan-type rooms must be fitted to
correctly-sized holes, correctly de-burred, and cleaned.
•
Use flat washers on both sides of the shield penetration.