Maintenance
| 11
C
LEANING
Removal of the tuner assembly must be
performed by qualified ETS-Lindgren
personnel.
•
Remove excess dust from the tuner and bearing assembly with a dry
cloth or vacuum.
•
Check the fixings to the shielded room and the tuner assemblies.
Doors
Periodically clean the brass or steel door frame and beryllium copper finger stock
with Channel Master® 9101 lubricant. See the decal applied to the door for more
detailed instructions. If staining or corrosion is present on the brass frame or
finger stock, polish lightly with 3M Scotch-Brite® pads. If extreme staining or
corrosion persists, call ETS-Lindgren for assistance.
Periodically inspect the roller cams and cam blocks to verify that a thin film of
lubricant is present. If additional lubricant is required, apply a very thin film of
silicone grease. The door maintenance kit included with each door contains the
necessary lubricants, cleaning material, and replacement finger stock.