Model 7600 Family PWC User’s Manual

7

Under Capacity Limits

Over Capacity Limits

Sealing

Internal Resolution

Dynamic Response

Communications

Under capacity indication will be given with
dashes appearing on the bottom line of the
display whenever the display is more than 10
division below the initial zero value.

Over capacity indication will be given with dashes
appearing in the upper line of the display when-
ever the weighed item exceeds 9 divisions over
the rated capacity of the unit. The scale will use
the initial zero value for reference for over
capacity determination.

Access to the calibration switch can be secured
with a lead wire or pressure sensitive security
seal. The remote and primary indicators have no
metrological features that require the use of a
security seal.

1 part in 2,000,000

The time interval of weight applied to scale until a
stable weight:

Transmitted

Displayed

0 - 1000d

500 msec

1200 msec

1000d+

750 msec

1400 msec

Maximum mean average

Factory default settings: 9600 baud, 7 data bits,
even parity, 1 stop bit.

Standard 9-pin pass through RS-232 interface
cable included, (not a null modem).

RS-232 bidirectional configurable 1200 to 19.2 K
baud. Transmits weight and scale status when-
ever ASCII “W” is sent by a remote device.