Calibration mode – Salter Brecknell 7680 User Manual

Page 23

background image

Model 7600 Family PWC User’s Manual

23

10. Press the UNITS key.

Parity selection is stored.

11. Press the ZERO key…

PROT is displayed.

12. Press the UNITS key…

The current setting is displayed. Use the
ZERO key to toggle between the four
choices: NCI STD, NCI SMA, AS350D,
PS6L.

13. Press the UNITS key…

Protocol selection is stored.

14. When finished configuring your scale, press

the ZERO key until the display shows DONE,
then press the UNITS key.

Or, move Switch 1 to CLOSED position for

normal weighing mode.

The calibration (CAL) mode menu lets you
calibrate your scale. The items in the calibration
menu are as follows:

Pounds/Kilograms (LB or 1000g)

Selects the unit of measure of your calibra-
tion test weights.

Scale or Classifier

When calibrating the scale for LB, you are
able to calibrate the unit as a scale or as a
classifier (weight classifier).

Calibration Mode

Calibration (CAL) Mode

Warning! Entering into
this mode can erase
the calibration already
saved. You need
approved calibration
weights to use
calibration mode.

Note: If this procedure
is attempted without
proper calibration
weights applied, the
scale will abort the
process and retain the
original calibration
data.