Model 7600 Family PWC User’s Manual

24

Unit On or Unit Off

When configured for UNIT ON, the scale will
allow you to switch between the selected
units of measure using the UNITS key.

Capacity (100, etc.)

Select the capacity of your scale.

Follow these steps to calibrate your scale. Refer
to Figure 2.

1.

From normal weighing mode, move Switch 1
to the Menu Mode or OPEN position…

DIAG is displayed. Press the ZERO key
until CAL is displayed. This stands for
calibration.

2.

Press the UNITS key to start calibration…

LBS or 1000g (kg) is displayed.

3.

Press the ZERO key to toggle between the
choices of units of measure (lb or kg). When
the choice you want is displayed, press the
UNITS key to accept…

If LBS was chosen for calibration, the
scale will display the current setting.
Press the ZERO key to toggle between
SCALE and CLASSIFIER. Calibrating as
SCALE = .5 division rounding. Calibrating
as CLASSIFIER = .9 division rounding.
Press UNITS key to accept.

4.

The current capacity is displayed. Press the
ZERO key to toggle between scale capacity
selections. Press the UNITS key to accept…

That choice is accepted and UNIT ON or
UNIT OFF is displayed.

Step -by-Step

Instructions for

CAL Mode