Diagnostics mode – Salter Brecknell 7680 User Manual
Page 18

Model 7600 Family PWC User’s Manual
18
The Diagnostic (Diag) mode menu lets you test
specific areas of the scale’s function.
These areas are:
Display (DISPLAY) – Shows the version and
revision of the software, followed by a display
segment test.
RAM (RA) – Performs a nondestructive test of
RAM in the processor. Displays PASS or FAIL.
ROM (RO) – Performs a checksum of all loca-
tions in ROM in the processor. Displays PASS or
FAIL.
Input/Output (I-O) – Data is output by the scale
and through the use of a loopback connector the
data is immediately read back into the receive
channel and verified against what was sent.
PASS or FAIL is displayed. Requires a jumper
(short) between transmit (Pin 2) and receive (Pin
3) data lines.
Division, test w/AZT (DIV-A) – Weight data is
normalized to 1,000,000 counts of displayed
resolution. AZT is enabled (Auto Zero Tracking).
Division, test w/o / AZT (DIV-N) – Weight data
is normalized to 1,000,000 counts of displayed
resolution. AZT is disabled.
Raw Counts (RA CNTS) – Non-normalized QDT
cell data (no zero tracking).
Diagnostics Mode
Diagnostics (DIAG )
Mode