Model 7800 Family Weight Classifiers User’s Manual
28
Re-Calibration Mode
Step-by-Step Instructions
for RE-CAL mode
Return to normal
operating mode
by pressing the
SW-1 switch
The re-calibration RE-CAL mode menu lets you
change the scale resolution (150lb / 75kg capaci-
ties only) or rounding method without using any
calibration weights. If you want to change the unit
of measure operation, you must perform a full
calibration using test weights.
For a scale originally calibrated in the lb. mode,
you may also change rounding methods (i.e.,
scale or classifier).
Follow these steps to re-configure your scale
(without weights). Refer to Figure 3.
1. From the normal weighing mode, move
Switch 1 to the Menu mode or OPEN posi-
tion...
DIAG is displayed.
2. Press the ZERO key until...
RE-CAL is displayed.
3. Press the TEST key...
ROUND is displayed.
To change the weight rounding method,
press the TEST key. The current round-
ing method is displayed.
4. Press the ZERO key to toggle between
SCALE and CLASS.
5. When the choice you want is displayed,
press the TEST key.
6. To change the capacity/resolution, press the
ZERO key until RESO is displayed.
7. Press the TEST key. The current capacity/
resolution setting is displayed.